Kamar uji iklim 350 liter Ex/lCTS
C-40/350 Ex/l -40°C /+180°C
Kamar uji iklim 350 liter Ex/l
CTS
C-40/350 Ex/l -40°C /+180°C
Kondisi
Bekas
Lokasi
Borken 

Gambar menunjukkan
Tampilkan peta
Data mesin
Harga & Lokasi
- Lokasi:
- Einsteinstraße 8a, 46325 Borken, Deutschland

Hubungi
Rincian penawaran
- ID iklan:
- A21151997
- Nomor referensi:
- 24669
- Pembaruan:
- terakhir pada 03.02.2026
Deskripsi
CTS
Model: C-40/350 Ex/l
Climate Test Chamber, Water-Cooled
ATEX Zone 1
The Ex/l designation indicates that this climate test chamber is suitable for use in potentially explosive atmospheres in Zone 1. This means it is intended for use in environments where explosive atmospheres may arise. Such equipment is specially designed and tested to ensure safety in these environments.
Temperature Testing:
Temperature range: -40°C to +180°C
Temperature deviation: ≤ ±0.3 K over time
Temperature change rate according to IEC 60068-3-5:
- Heating: 3.5 K/min
- Cooling: 4 K/min
Heat compensation:
- approx. 3500 W at +20°C
- approx. 2000 W at -20°C
Climate Testing:
Temperature range: +10°C to +90°C
Humidity range: 10% to 95% relative humidity
Dew point range: +5°C to 78°C
Temperature deviation: ≤ 1.0 K over time
Humidity deviation: ≤ +3.0 % r.h. over time
Technical data refer to an ambient temperature of +25°C, nominal voltage 400V, without test specimens, without heat compensation, and without additional equipment. Options may affect performance data.
Technical Specifications:
Bsdpfx Aeygui Sjf Sscu
Refrigeration unit: water-cooled, low-noise, fully hermetic, continuously variable capacity adjustment via electronic monitoring and control system, environmentally friendly refrigerant R404A
Test chamber: Floor load capacity 150 kg/m²
Insert grid: Stainless steel, supporting surface 620 x 600 mm, maximum load per grid 35 kg, maximum total grid load 100 kg
Control System:
32-bit controller
LCD operating panel
Digital display of set and actual values
Numerical input of set values
Plain text error messages
RS 232 interface
Ethernet interface
Free switching channels for customer-specific test specimen control
Dimensions:
Test volume: 350 liters
Internal dimensions: Height 750 mm × Width 650 mm × Depth 720 mm
External dimensions: Height 1800 mm × Width 850 mm × Depth 1640 mm
Operating Data:
Electrical connection: 400V +6/-10%, 3/N, 50 Hz
Nominal power: 7.15 kW
Nominal current: 14.3 A
Protection class: IP 54
Sound pressure level: [value missing]
Iklan ini diterjemahkan secara otomatis. Kesalahan terjemahan mungkin terjadi.
Model: C-40/350 Ex/l
Climate Test Chamber, Water-Cooled
ATEX Zone 1
The Ex/l designation indicates that this climate test chamber is suitable for use in potentially explosive atmospheres in Zone 1. This means it is intended for use in environments where explosive atmospheres may arise. Such equipment is specially designed and tested to ensure safety in these environments.
Temperature Testing:
Temperature range: -40°C to +180°C
Temperature deviation: ≤ ±0.3 K over time
Temperature change rate according to IEC 60068-3-5:
- Heating: 3.5 K/min
- Cooling: 4 K/min
Heat compensation:
- approx. 3500 W at +20°C
- approx. 2000 W at -20°C
Climate Testing:
Temperature range: +10°C to +90°C
Humidity range: 10% to 95% relative humidity
Dew point range: +5°C to 78°C
Temperature deviation: ≤ 1.0 K over time
Humidity deviation: ≤ +3.0 % r.h. over time
Technical data refer to an ambient temperature of +25°C, nominal voltage 400V, without test specimens, without heat compensation, and without additional equipment. Options may affect performance data.
Technical Specifications:
Bsdpfx Aeygui Sjf Sscu
Refrigeration unit: water-cooled, low-noise, fully hermetic, continuously variable capacity adjustment via electronic monitoring and control system, environmentally friendly refrigerant R404A
Test chamber: Floor load capacity 150 kg/m²
Insert grid: Stainless steel, supporting surface 620 x 600 mm, maximum load per grid 35 kg, maximum total grid load 100 kg
Control System:
32-bit controller
LCD operating panel
Digital display of set and actual values
Numerical input of set values
Plain text error messages
RS 232 interface
Ethernet interface
Free switching channels for customer-specific test specimen control
Dimensions:
Test volume: 350 liters
Internal dimensions: Height 750 mm × Width 650 mm × Depth 720 mm
External dimensions: Height 1800 mm × Width 850 mm × Depth 1640 mm
Operating Data:
Electrical connection: 400V +6/-10%, 3/N, 50 Hz
Nominal power: 7.15 kW
Nominal current: 14.3 A
Protection class: IP 54
Sound pressure level: [value missing]
Iklan ini diterjemahkan secara otomatis. Kesalahan terjemahan mungkin terjadi.
Penyedia
Catatan: Daftar gratis atau masuk, untuk mengakses semua informasi.
Terdaftar sejak: 2012
Kirim permintaan
Telepon & Faks
+49 2861 ... iklan
Iklan Anda telah berhasil dihapus
Terjadi kesalahan